MAV
Message Available Bit 4 (16)
Indicates that at least one reply message is waiting in the RPM4-AD IEEE-488 output
queue.
ERROR
Error Queue Not Empty Bit 2 (4)
Indicates that at least one command error message is waiting in the RPM4-AD
IEEE-488 error message queue. Use the
“ERR?”
query to get this message.
RSR
Ready Summary Bit 0 (1)
Indicates that an enabled bit in the Ready Status Register became set.
4.5.3
STANDARD EVENT REGISTER
The RPM4-AD contains an 8 bit Standard event register that reflects specific RPM4-AD
events. Enabled events in this register will set or clear the ESB bit of the Status Byte
Register.
Table 25.
8 bit standard event register
PON
(128)
URQ
(64)
CMD
(32)
EXE
(16)
DDE
(8)
QYE
(4)
RQC
(2)
OPC
(1)
This register can be read using the “*ESR?” query, Each of these status bits can set the ESB
bit of the Status Byte Register, causing a SRQ to occur IF the ESB bit is enabled to do so.
The Standard Event Status Enable Register (“*ESE” program message ) determines which of
these flags are able to assert the ESB bit. The description of these bits are given as:
PON
Power On Bit 7 (128)
Indicates that the RPM4-AD power has been cycled since the last time this bit
was read or cleared.
URQ
User Request Bit 6 (64)
Indicates that the RPM4-AD was set to local operation manually from the front
panel by the user (pressing the
[ESC]
key).
CMD
Command Error Bit 5 (32)
Indicates that a remote command error has occurred. A command error is
typically a syntax error in the use of a correct program message.
EXE
Execution Error Bit 4 (16)
Indicates if a remote program message cannot be processed due to device
related condition.
DDE
Device Dependent Error Bit 3 (8)
Indicates that an internal error has occurred in the RPM4-AD such as a transducer
time-out.
QYE
Query Error Bit 2 (4)
Indicates that an error has occurred in the protocol for program message
communications. This is typically caused by a program message being sent to
the RPM4-AD without reading a waiting reply.
RQC
Request Control Bit 1 (2)
This bit is not supported as the RPM4-AD cannot become the active controller in
charge.
OPC
Operation Complete Bit 0 (1)
Indicates that the RPM4-AD has completed all requested functions.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...