3. OPERATION
Page 37
© 2005-2007 DH Instruments, a Fluke Company
Pressing [ENT] while in the Average DISPLAY aborts the current averaging
period and causes a new one to begin. [ENT] can thus be use to trigger a
new averaging period on demand.
In the Average DISPLAY the Ready/Not Ready indication applies to the result
of the previous averaging period (see Section 3.2.4). <Ready> indicates
that all readings during the previous averaging period met the stability
criterion. <Not Ready> indicates that one or more readings were outside of
the stability criterion.
Changing the pressure unit of measure, measurement mode or range while
the averaging screen is active, starts a new averaging period.
To go to a DISPLAY other than Average, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.2
Rate
PURPOSE
To activate the Rate DISPLAY.
See
Section
3.3.6 [DISPLAY],
Principle.
OPERATION
To activate the Rate DISPLAY press
[DISPLAY]
,
.
The display is:
Select whether the rate should be
displayed in unit per minute or unit per
second. The default is per minute.
Display rate per:
1minute 2second
When the selection is made, operation returns to the MAIN RUN screen with the
Rate DISPLAY active.
With the Rate DISPLAY active the MAIN RUN screen is:
1. Current rate of change of in current unit of
measure per minute (
</m>
) or second (
</s>
).
The default is per minute.
20473.6 ft a HL
R 378.3 /m
The Rate DISPLAY is different and separate from the stability setting which
is used to set the stability criterion on which the Ready/Not Ready
indication is based (see Sections 3.4.3 and 3.2.4). The Rate DISPLAY only
causes the current rate of change to be displayed and has NO affect on the
stability setting or the Ready/Not Ready condition.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...