5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS
Page 119
© 2005-2007 DH Instruments, a Fluke Company
c) Set full scale absolute pressure.
d) Dwell for five minutes.
Repeat the exercising sequence twice then proceed with the calibration
sequence defined in Table 28 (pressure is already at the first point).
All pressure points are absolute pressure.
A dwell time of at least 60 seconds after setting the pressure is recommended
before taking data at each point.
Table 28.
Calibration point sequence, A160K Q-RPT
CALIBRATION
SEGMENT
POINT
NO.
POINT
[kPA (PSI)]
1
160 (23.2)
2
120 (17.4)
3
80 (11.6)
4
40 (5.8)
Descending
5
6 (0.9)
6
40 (5.8)
7
80 (11.6)
8
120 (17.4)
Ascending
9
160 (23.2)
It is not required that the calibration pressure standard apply precisely the nominal
pressure value of each calibration point as long as the exact value of the applied pressure
is known. Best results are obtained if the actual applied pressure is within 2 % of the
recommended nominal point.
5.2.4.2
A350K Q-RPT CALIBRATION SEQUENCE
The calibration sequence starts with exercising of the Q-RPT:
a) Set full scale absolute pressure.
b) Dwell for five minutes.
c) Apply vacuum (very low absolute pressure)
d) Dwell for five minutes.
Repeat the exercising sequence twice then proceed with the calibration
sequence defined in Table 29.
All pressure points are absolute pressure.
A dwell time of at least 60 seconds after setting the pressure is recommended
before taking data at each point.
Table 29.
Calibration point sequence, A350K Q-RPT
CALIBRATION
SEGMENT
POINT
NO.
POINT
[kPA (PSI)]
1 Lowest
pressure
1
2
87.5 (12.7)
3
175 (25.4)
4
262.5 (38.1)
Descending
5
350 (50.8)
6
262.5 (38.1)
7
175 (25.4)
8
87.5 (12.7)
Ascending
9 Lowest
pressure
1
1. 1 kPa (0.15 psi) or the lowest point that can be set reliably by
the reference piston gauge .
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...