RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
© 2005-2007 DH Instruments, a Fluke Company
Page 10
2.4
POWER-UP AND VERIFICATION
2.4.1
SWITCH POWER ON
Actuate the power switch on the RPM4-AD rear panel (if a 12 VDC power supply is already
connected to the
12 VDC
connection, RPM4-AD power is already on). Observe the front panel
display as RPM4-AD initializes, error checks and goes to the MAIN RUN screen (see 3.1.1).
If the RPM4-AD fails to reach the MAIN RUN screen, service is required. Record the
sequence of operations and displays observed.
SDS is CLOSED at power up and the TEST ports are shut off. This causes <SDS
CLOSED> to be flash periodically in the MAIN RUN screen in place of the measured
pressure value.
2.4.2
CHECK PRESSURE MEASUREMENT OPERATION
2.4.2.1
CHECKING ABSOLUTE MODE PRESSURE MEASUREMENT
Check the RPM4-AD operates properly in
absolute
mode.
Make sure that the
TEST (Ps)
and
TEST (Pt)
ports are vented to atmosphere.
Press the
[MODE]
function key and select
<absolute>
mode (see Section 3.3.3).
Use
[UNIT]
to select a
pressure
unit of measure (not altitude) (see Section 3.3.2).
If SDS is CLOSED (
<SDS CLOSED>
flashes over the display of pressure on
the top line of the display), Press
[SDS]
,
<2yes>
to OPEN SDS (see Section 3.3.8).
Do NOT OPEN SDS with a pressure higher than the maximum pressure of the
Q-RPT applied to the TEST port. Damage to the Q-RPT may result.
The RPM4-AD should be measuring the current value of atmospheric pressure.
Check that the value agrees with the local value of atmospheric pressure within
measurement tolerances (see Section 1.2.2.1). If RPM4-AD does not agree
within tolerance, it may need to be AutoZeroed (see Section 3.3.9), calibrated
(see Section 5.2) or repaired.
2.4.2.2
CHECKING GAUGE MODE PRESSURE MEASUREMENT
Check the RPM4-AD operates properly in
gauge
mode.
Make sure that the
TEST (Ps)
and
TEST (Pt)
ports are vented to atmosphere.
Press the
[MODE]
function key and select
<gauge>
mode. Use
[UNIT]
to select a
pressure
unit of measure (not airspeed) (see Section 3.3.2).
If SDS is CLOSED (
<SDS CLOSED>
flashes over the display of pressure on the
top line of the display), OPEN SDS. Press
[SDS]
,
<2yes>
to OPEN SDS.
Do NOT OPEN SDS with a pressure higher than the maximum pressure of the
Q-RPT applied to the TEST port. Damage to the Q-RPT may result.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...