![Fluke RPM4-AD Скачать руководство пользователя страница 137](http://html1.mh-extra.com/html/fluke/rpm4-ad/rpm4-ad_operation-and-maintenance-manual_2310054137.webp)
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS
Page 127
© 2005-2007 DH Instruments, a Fluke Company
5.6.2
RPM4-AD A350K/A160K (FIXED WING)
Table 31.
Q-RPT module valve states, RPM4-AD A350K/A160K
LO Q-RPT ACTIVE
ABSOLUTE AND GAUGE MODES
(altitude, no airspeed)
HI Q-RPT ACTIVE
ABSOLUTE AND GAUGE MODES
(airspeed, at ambient pressure
altitude only)
HI Q-RPT ACTIVE
DIFFERENTIAL MODE
(airspeed and any altitude)
VALVE STATE
CONDITION
A
B
C
D
E
SDS Open
O
C O
C
C
SDS Closed
C C O O C
Power off
C C O O C
VALVE STATE
CONDITION
A
B
C
D
E
SDS Open
C
O
C
O C
SDS Closed
C C O O C
Power off
C C O O C
VALVE STATE
CONDITION
A
B
C
D
E
SDS Open
O
O
C
C
C
SDS Closed
C C O O C
Run AutoZ
O
C
C
C
O
Figure 9.
Pneumatic schematic and valve states, RPM4-AD A350K/A160K (fixed wing)
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...