RPM4-AD™ OPERATION AND MAINTENANCE MANUAL
© 2005-2007 DH Instruments, a Fluke Company
Page 40
When the RPT DISPLAY is active and the unit of measure for the active Q-RPT
IS an airspeed unit, the MAIN RUN screen is:
1.
Active Q-RPT measurement.
2.
Active Q-RPT indicator.
3.
Altitude at which airspeed is being measured.
498.778 kts d zHd
*22477.5 ft a zLo
When the RPT DISPLAY is active and the unit of measure for the active Q-RPT
is NOT an airspeed unit, the MAIN RUN screen is:
1.
Active Q-RPT measurement.
2.
Active Q-RPT indicator.
3.
Inactive Q-RPT measurement.
150.582 kPa a zHi
* 98.477 kPa a zLo
If SDS Self Defense System (see Section 3.2.8) is CLOSED on the inactive Q-
RPT when the RPT DISPLAY is activated, SDS can be OPENED on the inactive Q-
RPT using [SPECIAL], <4SDS>, <1temp open/close> (see Section 3.5.4). In
differential measurement mode, SDS is be opened simultaneously on both Q-
RPTs using [SDS] (see Section 3.3.8).
The RPT DISPLAY is the default display in differential measurement mode.
When the RPT display is active, executing a Q-RPT change to the inactive
Q-RPT makes the inactive Q-RPT the active Q-RPT. The DISPLAY defaults
back to Rate ( Section 3.3.6.2).
To go to a DISPLAY other than RPT, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.5
Hi/Lo
PURPOSE
To activate the Hi/Lo DISPLAY.
See
Section
3.3.6,
Principle..
OPERATION
To activate the Hi/Lo DISPLAY press
[DISPLAY]
,
<5Hi/Lo>
. Pressing
<5Hi/Lo>
resets the Hi/Lo values and returns to the MAIN RUN screen with the Hi/Lo
DISPLAY active.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...