4. REMOTE OPERATION
Page 93
© 2005-2007 DH Instruments, a Fluke Company
L2 / L3
Purpose
Read or set the type of program command format to use (classic or enhanced).
Command
“L2” enables “classic” mode
“L3” enables “enhanced” mode
Defaults “L2”
(Classic
mode)
Remarks
The user can select the type of remote command format using these simplified commands.
There is no query format. This format must agree with the format sent to the RPM4-AD. This
command is a replacement for the “MSGFMT” command.
Example
(enhanced)
Cmd sent:
Query reply:
“L3”
“L3”
(No reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“L2”
“L2”
See Also
“MSGFMT”
4.3, 3.5.2.3
LL
n
Purpose
Read or set the lower pressure limit for the active range and measurement mode (negative
gauge or differential measurement mode only).
Command “LL
n
limit
”
Classic “LL
n
=
limit
”
Optional Suffix
“n”
The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT if the HL RPT is not active
Specify the HL Q-RPT if the HL Q-RPT is active
‘2’ Specify the Lo Q-RPT if the HL Q-RPT is not active and the
measurement mode is not differential
‘3’ Specify the HL Q-RPT if the HL Q-RPT is active
Arguments
"limit
”
:
The lower pressure limit for the current pressure range in the specified Q-
RPT’s current unit of measure unit and measurement mode. Value is always
a negative value of gauge pressure.
Remarks
Negative gauge and differential modes in RPM4-AD each have a separate lower limit. If the
pressure is less than the lower limit, the pressure display flashes. This feature can used to
prevent accidental over (under) pressure of a device under test.
Example
(enhanced)
Cmd sent:
Reply:
“LL -4”
“-4 kPa g”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“LL? -4”
“-4 kPa g”
Example
(classic)
Sent:
Reply:
“LL=-4”
“-4 kPa g”
Errors ERR#
6:
ERR# 10
ERR# 23
The
limit
argument is invalid.
The suffix is invalid
The mode must be negative gauge to specify a lower limit.
See Also
“UL”, “MMODE”
3.4.4, 3.3.3
LOCAL
Purpose
Returns control to the RPM4-AD front panel.
Command “LOCAL”
Classic “LOCAL”
Remark
The REMOTE program message can lock the front panel out completely. The user can return
to local operation by sending the LOCAL program message, sending the IEEE-488 ‘GTL’
command (if in enhanced format), or by cycling RPM4-AD power.
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL”
“LOCAL”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL?”
“LOCAL”
Example
(classic)
Sent:
Reply:
“LOCAL”
“LOCAL”
See Also
“REMOTE”
MEM
Purpose
Read the status from the power-up memory test.
Query “MEM?”
Classic “MEM”
Remarks
The RPM4-AD system memory stores the user settings (units, resolution) and retains them
when the unit is OFF. On power-up, this memory is checked. If this memory is corrupted, all
user settings are reset to default , and the MEM status is set to reflect this.
Example
(enhanced)
Query sent:
Reply:
“MEM?”
“0” RPM4-AD data corrupted and was set to factory defaults.
“1” The memory was found to be OK on power-up.
Example
(classic)
Sent:
Reply:
“MEM”
“MEM=0” RPM4-AD data corrupted and was set to factory defaults.
“MEM=1” The memory was found to be OK on power-up.
See Also
3.5.9,
Error! Reference source not found.
Содержание RPM4-AD
Страница 10: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 22: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...
Страница 90: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 80 N NO OT TE ES S...
Страница 138: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 128 N NO OT TE ES S...
Страница 142: ...RPM4 AD OPERATION AND MAINTENANCE MANUAL 2005 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S...