The
P2131
FIRMWARE
UPDATE
KIT
includes
the
following
items:
‐
CD
with
Cscape
programming
software
‐
CD
with
the
new
PLC
firmware
‐
Micro
SD
card
and
adapter
with
the
sterilizer
software
program
‐
Micro
SD
extender
cable
‐
Communication
cable
‐
Instruction
sheet
Step
1
–
Loading
the
New
Firmware
1.
Insert
the
CD
with
Cscape
software
and
install
the
Cscape
programming
software
to
your
computer.
NOTE:
Computer
must
be
a
PC
with
Windows
7
or
newer
operating
system.
2.
Insert
the
CD
with
the
new
PLC
firmware,
and
run
the
XLe
‐
XLt
‐
firmware
application
from
the
CD.
This
will
save
several
files
into
the
Cscape/Firmware
folder
on
your
computer.
Verify
that
these
files
are
in
the
correct
folder
on
your
computer.
3.
Ensure
the
sterilizer
is
OFF,
and
connect
the
communication
cable
to
a
serial
COM
port
on
your
computer
and
to
MJ1
on
the
Programmable
Logic
Controller
(PLC).
4.
Turn
the
sterilizer
ON.
5.
Open
Cscape
on
your
computer
and
establish
communication
between
Cscape
and
the
PLC.
a.
At
the
ABOUT
Cscape
window,
choose
OK.
b.
On
the
CONNECTION
WIZARD
window,
select
SERIAL
and
choose
NEXT.
c.
On
the
following
window,
enter
the
COM
port
number
that
you
are
using
for
communication
and
choose
FINISH.
6.
Within
Cscape
select
FILE
and
then
NEW
to
create
a
new
blank
program.
This
blank
program
is
only
used
to
access
the
firmware
update
wizard.
7.
Start
the
firmware
update
by
selecting
FILE
and
then
FIRMWARE
UPDATE
WIZARD.
8.
In
the
Firmware
Update
Wizard,
select
XLT
for
the
Product
Type
and
CsCAN
NETWORK
for
Networking
and
choose
OK.
9.
On
the
following
window
choose
OK
to
accept
FIRMWARE
UPDATES
USE
HIGH
SPEED
SERIAL
COMMUNICATIONS.
10.
The
PLC
will
enter
the
System
Firmware
Loader
mode
and
the
FIRMWARE
UPDATE
window
will
appear.
Choose
START
to
begin
the
update.
CAUTION
‐
Once
the
download
starts,
do
not
interrupt
the
communication
link
while
the
download
is
in
progress.
Doing
so
will
permanently
lock
the
PLC.
11.
Wait
for
the
firmware
update
to
complete.
You
will
see
a
progress
bar
as
the
firmware
is
being
updated.
12.
When
you
see
DOWNLOAD
COMPLETE,
choose
CLOSE.
13.
At
the
UPDATE
COMPLETE
window,
choose
OK.
WARNING:
Firmware
updates
typically
delete
the
user
applications
to
ensure
compatibility.
After
updating
the
firmware,
you
will
need
to
reload
the
sterilizer
software
program.
Make
sure
the
sterilizer
software
program
is
available
before
beginning
the
firmware
update
.
93
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...