FDA
‐
REQUIRED
WATER
CHANGEOUT
EVERY
14
DAYS
OR
100
CYCLES
(WHICHEVER
OCCURS
FIRST)
CHECK
TASK
1.
Clean
the
SMU
chamber
thoroughly
using
a
mild
detergent:
a.
Disconnect
the
condensate
hose
(red
‐
tagged)
at
the
WRS
and
place
the
end
into
a
bucket.
b.
Drain
any
cleaning
water
/
mild
detergent
into
the
bucket.
c.
Flush
the
hose
with
fresh
water
and
then
reconnect
the
hose
to
the
WRS.
WARNING:
Do
not
allow
any
water
or
detergent
to
drain
into
the
WRS.
2.
Drain
all
the
water
from
the
SMU
jacket
by
following
the
Jacket
Blowdown
Procedure
in
Section
4.1.2.
WARNING
:
To
avoid
a
potential
burn
hazard,
ensure
the
high
‐
temperature
drain
hose
is
used.
3.
Drain
the
WRS
by
attaching
the
drain
hose
and
opening
the
drain
valve.
4.
SWS
maintenance:
a.
Test
the
TDS
level
of
the
water
coming
out
of
SWS
and
replace
the
deionized
water
cartridge
as
needed.
Reference
Section
3.4.1
Testing
Total
Dissolved
Solids
and
Section
4.3.4
SWS
Cartridge
Replacement.
b.
Test
the
chlorine
level
of
the
water
coming
out
of
the
SWS
Reference
Section
3.4.3
Testing
Free
Chlorine
Level.
Water
can
foul
and
become
discolored
if
not
used
regularly.
Before
each
use,
flush
SWS
until
water
is
clear
(2
minutes).
5.
Refill
the
entire
system
with
fresh
potable
water
(Reference
Section
2.5
–
Water
Supply
Quality).
6.
Run
a
VACUUM
LEAK
TEST.
54
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...