ALARM
CODE
POSSIBLE
CAUSES
CORRECTIVE
ACTION
06
Insufficient
Vacuum
Loose
vacuum/exhaust
hose
fittings
(white
‐
tagged
hose)
Pump
not
operating
in
WRS
Hot
water
in
WRS
Eductor
problem
in
WRS
Vacuum
valve
not
open
far
enough
Inspect
O
‐
rings
at
each
end
and
re
‐
tighten
the
hose.
Check
WRS
water
level,
power
cord,
and
power
switch.
Ensure
both
WRS
fans
are
running
when
pump
is
ON.
Ensure
water
level
is
at
least
1/2
full.
Too
little
water
will
tend
to
overheat
and
cause
low
eductor
performance.
Contact
a
qualified
maintenance
technician
to
inspect
eductor
for
leak
or
worn/broken
nozzle
inside
eductor.
Replace
if
needed.
Contact
a
qualified
maintenance
technician
to
inspect
solenoid
valve
and
replace
if
needed.
NOTE:
Reference
Appendix
A.5
Vacuum
Leak
Test
Troubleshooting
Guide
for
more
help
on
troubleshooting
a
potential
leak
in
the
system.
07
Chamber
Pressure
High
Jacket
pressure
high
Reference
alarm
code
09
Jacket
Pressure
High.
63
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...