39.
Ensure
the
jacket
vent
valve
on
the
back
of
the
SMU
is
fully
open
and
close
the
chamber
vent
valve.
This
will
allow
the
jacket
to
vent
the
air
while
it
is
filled
with
water
from
the
WRS.
Minimizing
air
in
the
jacket
improves
sterilizer
performance
since
air
acts
as
an
insulator
during
sterilization.
40.
Press
and
hold
the
“Water
to
Sterilizer”
button
on
the
WRS
and
fill
the
SMU
to
the
HIGH
mark
on
the
level
sight
glass.
This
should
pump
about
4.25
gallons
of
water
into
the
jacket
of
the
sterilizer.
Add
water
to
the
WRS
with
the
SWS
as
needed
to
fill
the
SMU
to
the
HIGH
mark.
CAUTION
:
Verify
the
SMU
sight
glass
fills
from
the
bottom.
If
not,
contact
a
qualified
maintenance
technician
to
inspect
the
sight
glass
plumbing
for
debris.
CAUTION
:
Never
add
water
to
the
jacket
during
a
cycle.
This
will
cause
the
jacket
to
cool,
lose
pressure,
and
alarm.
41.
Firmly
close
the
jacket
vent
valve
on
the
back
of
the
SMU.
42.
Refill
the
WRS
back
to
its
HIGH
mark
and
disconnect
the
SWS
hose.
43.
Reset
the
“02
–
LOW
WATER
LEVEL”
alarm.
If
the
alarm
clears,
the
liquid
level
switch
is
functioning
correctly
in
the
“Water
OK”
position.
CAUTION
:
If
this
“Low
Water”
alarm
does
not
reset,
contact
a
qualified
maintenance
technician
to
inspect
the
liquid
level
switch
and
liquid
level
switch
plumbing.
44.
On
the
"CONFIRM
SETUP”
screen,
make
any
necessary
updates
to
the
TIME,
DATE,
or
UNIT
ID.
To
do
so,
touch
the
box
to
the
right
of
the
metric
that
needs
to
be
updated.
A
new
screen
will
appear
for
the
user
to
make
any
desired
updates.
This
screen
will
appear
every
time
the
Control
Power
is
turned
ON.
NOTE
:
The
correct
water
levels
are
HIGH/HIGH
‐
when
the
water
is
at
the
HIGH
mark
in
the
jacket,
the
water
level
should
be
at
or
near
the
HIGH
mark
in
the
WRS.
26
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...