16.
Remove
the
lower
front
panel
of
the
SMU
to
access
the
connection
for
the
condensate
hose.
Use
the
multi
‐
tool
provided
in
the
SAB
to
loosen
the
front
panel
latches,
and
then
gently
remove
the
panel.
17.
On
the
bottom
left
side
of
the
SMU
is
an
access
hole
for
the
condensate
hose
(red
‐
tagged).
Push
the
hose
through
this
hole
and
unscrew
the
cap
on
the
left
side
of
the
drain
manifold.
Ensure
the
O
‐
ring
on
the
manifold
fitting
is
in
place,
and
then
connect
the
hose
to
the
condensate
outlet
on
the
chamber
drain
manifold.
Hand
‐
tighten
the
nut
and
then
tighten
45°
(max)
using
the
multi
‐
tool.
NOTE:
If
the
fitting
O
‐
ring
is
missing,
spare
O
‐
rings
are
provided
in
the
SAB.
16