1. Cap
2. Stainless Steel Tube with silicone coating
3. Access vent
4. Unique Identification Number
5. Counter Arm
6. Number Scale
7. Tracker Chip
VERIFY Bowie Dick Indicator Strips:
2021/01
116
Insert into Pack
1
VERIFY™
All-In-One BD STRIP
Dark bar must
cover window.
The reusable test packs use the VERIFY Bowie Dick Indicator Strips to challenge the air removal efficacy of prevacuum steam steril-
izers operating at 270°F. Either the sterilizers preprogrammed Daily Air Removal Test Cycle (Bowie Dick Test Cycle) or a prevacuum
cycle operating at 134°C/273°F with a 3.5 minute exposure and a 2 minute dry time may be used.
During the test cycle, the steam sensitive chemical located within the strip will wick across the viewing window. The dark color will
completely fill the window if all air was removed from the reusable test pack and replaced by steam.
The VERIFY Reusable test pack comprises three models: (1) Standard Test Pack, (2) Test Pack with Step Counter, (3) Test Pack with
Tracker Tag.
VERIFY All-In-One STEAM
Reusable Test Pack (Standard)
VERIFY All-In-One STEAM
Reusable Test Pack with Step
Counter
VERIFY All-In-One STEAM
Reusable Test Pack with Tracker
Tag
2021/01
116
Insert into Pack
1
VERIFY™
All-In-One BD STRIP
Dark bar must
cover window.
UNPROCESSED
2021/01
116
Insert into Pack
1
VERIFY™
All-In-One BD STRIP
Dark bar must
cover window.
PROCESSED REJECT
2021/01
116
Insert into Pack
1
VERIFY™
All-In-One BD STRIP
Dark bar must
cover window.
PROCESSED ACCEPT
COMPONENTS
DARK BAR MUST COMPLETELY COVER WINDOW
137
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...