WRS
K4
Relay
and
Liquid
Level
Switch
Check
CAUTION:
High
Voltage
contacts
exposed
with
the
WRS
Top
Panel
removed.
Electrical
shock
hazard
when
using
a
multi
‐
meter
to
check
the
K4
Relay
and
Liquid
Level
Switch.
CHECK
TASK
1.
Remove
the
Top
Panel
from
the
WRS.
2.
Connect
power
to
the
WRS
from
the
SMU,
but
do
not
slide
the
WRS
underneath
the
SMU
or
connect
the
white
hoses.
3.
Turn
the
SMU
and
WRS
on.
With
the
WRS
empty,
use
a
multi
‐
meter
to
check
terminal
#10
on
S/N’s
AFS
‐
0001
through
AFS
‐
0236
or
terminal
#14
on
S/N’s
after
AFS
‐
0236
for
115
VAC.
4.
With
the
WRS
empty,
use
a
multi
‐
meter
to
check
for
no
continuity
across
terminal
#1
and
#5
on
K4
Relay.
5.
Fill
the
WRS.
Use
a
multi
‐
meter
to
check
terminal
#10
on
S/N’s
AFS
‐
0001
–
0236
and
terminal
#14
on
S/N’s
after
AFS
‐
0236
for
0
to
40
VAC.
6.
With
the
WRS
filled,
use
a
multimeter
to
check
for
continuity
across
terminal
#1
and
#5
on
K4
Relay.
Inspect
SMU
Fill
Check
Valve
&
WRS
Fill
Valve
CAUTION:
Burn
Hazard.
Hot
water
and/or
steam
can
remain
in
the
hose
that
will
be
removed
below
when
it
is
disconnected.
CHECK
TASK
1.
Ensure
the
Sterilizer
is
fully
setup
and
heated
‐
up.
2.
Remove
the
Recovered
Water
from
WRS
Hose
(Blue
‐
tagged)
from
the
SMU
and
WRS.
CAUTION:
Do
not
remove
any
other
hoses.
Check
to
ensure
that
no
water
is
exiting
the
SMU
Blue
‐
tagged
VCO
fitting.
3.
Start
an
IUSS
cycle,
which
will
start
the
pump
on
the
WRS.
Check
to
ensure
that
no
water
drips
from
the
WRS
Blue
‐
tagged
VCO
fitting.
4.
Abort
the
cycle
and
allow
the
SMU
to
vent.
5.
Verify
that
O
‐
rings
are
in
place
and
reconnect
the
hose.
79
Содержание P2131
Страница 121: ... THIS PAGE INTENTIONALLY LEFT BLANK 115 ...
Страница 130: ...SECTION A APPENDIX 124 ...
Страница 131: ...A 1 Swagelok Fitting Installation Instructions 125 ...
Страница 135: ...A 2 PLC Data Sheet 129 ...
Страница 141: ...A 3 VERIFY Product Information 135 ...
Страница 147: ...A 4 WRS Pump Information 141 ...
Страница 158: ...A 5 Vacuum Leak Test Troubleshooting Guide 152 ...