3. Description
DEPOLOX® 5 E flow cell
EN
3.3
Spare Parts/extensions
3.4
Optional accessories
Please find order numbers in chapter 6.
•
Strainer (chapter 4.8)
•
Booster pump (chapter 4.8)
•
Pressure reducing valve (chapter 4.8)
•
PVC- or PE hose and hose connectors
(chapter 4.8)
•
CAN bus extension cable (chapter 4.4)
3.5
Design and functions
3.5.1 Overall design
Abb. 1 DEPOLOX® 5 E flow cell, non-pressurized
A Cover for holding the sensors and LED
glow stick (cell body cover)
B
Plastic housing with removable housing
cover
C
Cell body
D DFMe electronics module
E
Calibration holding clip
F
3-electrode measuring cell with electrolyte
tank
G Cap of the 3-electrode measuring cell
H Sample water outlet, in pressurized versi-
on also with ball valve
I
Sample taking unit (drain)
J
Sample water inlet
K
Shut-off ball valve
L
Filter unit
M Check valve housing
N Flow control valve
O Multi-sensor
P
Module SiDiSens conductivity (optional)
Part No.
Designation
W3T320090*
Sensor measuring module
DFMe Total chlorine TC3 CAN
W3T417351*
Cell body cover, fully pressu-
rized version
W3T320088
Sensor measuring module
DFMe pH
W3T320089
Sensor measuring module
DFMe ORP (Redox)
W3T320091
Sensor measuring module
DFMe Conductivity LF325 -
60 mS/cm
W3T320092
Sensor measuring module
DFMe Conductivity LF325 -
600 µS/cm
NOTICE
*The pressurized version can be retrofitted
with the total chlorine membrane sensor TC3
CAN. Please note that the standard cell body
cover must be replaced by an optional cell
body cover (part no. W3T417351), see chap-
ter 4.21.2).