29
DEPOLOX® 5 E flow cell
4. Installation
EN
4.13.2 Pressurized version
1 Extract the knurled nut.
2 Remove the long transport cap from the
electrolyte tank and replace with the
short operating cap (enclosed).
3 Fit the knurled nut.
Fig. 9
Detail: cell body cover, pressurized
version
A Transport cap (long)
B
Stopper (short)
4.14 Filling electrode cleaning sand
The electrode cleaning sand (W3T158743) is
supplied in a plastic bottle, the cap of which
serves as a measuring beaker.
1 Close the shut-off ball valve on the sample
water inlet.
2 On the pressurized version, close the
shut-off ball valve on the sample water
outlet.
3 On the non-pressurized version, take out
a plug or sensor from the cell body cover.
4 On the pressurized version, remove the
protection cap or sensor on the cell body
cover.
5 Fill half a cap from the plastic bottle with
cleaning sand and pour it into the cell
body through a sensor mounting hole
(approx. 1/3 cm³ of cell sand).
6 Make sure that the opening and the
thread are clean; if necessary rinse with
distilled water.
7 Insert the plug or a sensor in the cell body
cover.
8 Open the shut-off ball valve at the sample
water inlet.
9 On the pressurized version, open the
shut-off ball valve on the sample water
outlet.
ATTENTION
During operation, the cap must always be in
place.
A
B