3. Description
DEPOLOX® 5 E flow cell
EN
3.
Description
The DEPOLOX® 5 E flow cell (module type D01)
is part of the Pool Management System DEPO-
LOX® 5 E 700 P and the measuring and control
unit DEPOLOX®
5 E 700 M.
The DEPOLOX® 5 E flow cell (module type D01)
is a special flow cell adapter for sample water.
In combination with the 700 P electronics
module (module type E01) or 700 M electro-
nics module (module type E01) and the integ-
rated sensors, the DEPOLOX® 5 E flow cell
measures and controls parameters free chlo-
rine, pH value, redox voltage and temperature.
To determine parameters total chlorine, com-
bined chlorine and conductivity, further sen-
sors can e installed.
DEPOLOX® 5 E flow cell is available in both a
non-pressurized and a pressurized version.
The two versions differ in the number of sen-
sors that can be installed and in the design of
the sample water outlet. The non-pressurized
version is characterized by a free sample water
outlet. With the pressurized version, the sam-
ple water is returned into the system’s circuit.
The for sensors required for the measurement
tasks (optional) and the LED glow stick are
inserted or screwed into the mounting hole in
the cell body cover.
The maintenance-free flow control valve is the
main hydraulic part of DEPOLOX® 5 E flow cell.
It has the task of keeping the sample water
flow constant, irrespective of fluctuations in
operating pressures.
The multi-sensor monitors the correct flow
and records the temperature of the sample
water. The metallic sensor housing guarantees
good sample water earthing. The continuous
hydromechanical cleaning of the electrode of
the three-electrode measuring cell effectively
prevents natural contamination of the elec-
trode surfaces and ensures long-term stable
chlorine measurement.
DFMe electronics module converts the analog
sensor signals for digital transmission to 700 P
electronics module or 700 M electronics
module. If a membrane sensor for total chlo-
rine TC3 CAN (optional) is used, this is connec-
ted directly to DFMe electronics module or
routed through the CAN connection on con-
ductivity module SiDiSens (optional) to DFMe
electronics module. The conductivity sensor
(optional) is connected to DFMe electronics
module through conductivity module SiDi-
Sens.
3.1
Versions
The DEPOLOX® 5 E flow cell (module type D01)
can be comined with 700 P electronics module
(module type E01) or 700 M electronics
module (module type E01). The following ver-
sions are available:
•
Pool Management System DEPOLOX® 5 E
700 P
•
Measuring and control unit DEPOLOX® 5 E
700 M
3.1.1 Pool Management System
DEPOLOX® 5 E 700 P
The DEPOLOX® 5 E flow cell (module type
D01) is factory-configured according to the
customer’s specific requirements with the 700
P electronics module (module type E01) accor-
ding to the variant code. The scope of delivery
differs in the versions of DEPOLOX® 5 E flow
cell, the sensors, electronics components and
accessories. The variant code consists of 12
digits and is printed on both the type plate and
packaging sticker.
Part no.
DEPOLOX® 5 E flow cell
W3T350215 Non-pressurized version
W3T350216 Pressurized version