background image

6. Spare parts, Accessories and Retrofit kits

DEPOLOX® 5 E flow cell

42

EN

6.

Spare parts, Accessories and Retrofit kits

6.1

DEPOLOX® 5 E flow cell (module type D01) – non-pressurized version 
W3T350215

NOTICE

For reasons of safety, only use original spare parts. Please contact our customer service if you
need any spare parts.

Item

Part No.

Designation

1 W3T247776 Main 

housing

3 W3T247777 Housing 

cover

7

W3T166170 Shut-off 

valve

8 W3T158593 

Discharge 

nozzle

9 W2T507615 Flat 

nut

10 W3T158561 

Cell 

body

11 W3T164226 

Compression 

spring

12 W3T158569 

Membrane 

unit

13 W3T160654 

O-ring

14 

W3T158595  Control valve body

15 W2T504209 

Plastic 

cartridge

16 

W3T160649  V profile clamp

17 W3T158567 

Square 

nut

18 W2T504659 

Cheese-head 

screw

19 W2T506019 

Washer

20 W3T158572 

Valve 

pin

21 W3T172795 

Compression 

spring

22 W3T158573 

Adjusting 

screw

23 W3T160357 

O-ring

26 

W3T160648  Check valve housing

28 W3T161396 

O-ring

29 

W3T169827  Float with magnet

30 W3T172946 

Ball

31 W3T172949 

O-ring

32 W3T159707 

Insert

30-33 W3T163739 Spherical set cpl.

29-33 W3T158603

Check valve unit cpl. 
with float cone

33 W3T172975 

O-ring

35 W3T158602 

Filter 

housing

33-34 W3T166194

Fine filter cpl. with O-
ring

36 W2T505463 

Panh. 

tension 

screw

39 W3T172041 

Securing 

ring

40 

W3T158576 Outlet drain pipe

41 W3T172997 

O-ring

42 W3T164597 

O-ring

43 W3T158575 

Drain 

screw

44 W3T166160 

EPDM 

flat 

gasket

45 W3T172556 

O-ring

39-45 W3T166171 Sample taking unit
46 

W3T320060 Cell body cover

47 W3T160657 

O-ring

48 W3T165266 

Knurled 

nut

50 W3T172861 

O-ring

51 W2T863568 

Adapter

54

W3T438413 Hose

55 W2T505093

Angle-reducing con-
nector

56 W3T166209 Electrode 

mount

57 W3T168875 

O-ring

58 W3T163795 

Working 

electrode

59 W3T167461 

Counter 

electrode

60 W3T168904 

O-ring

56-60 W3T166210

Electrode mount cpl. 
with electrode

61 W3T158562 

Sealing 

cap

62 W3T168868 

O-ring

63 W2T504177 

Cable 

union

64 W3T160549 Hex 

nut

66 W3T159653 

Electrode 

housing

67-69 W3T159870 Diaphragm compl.
70 W3T169295 

Reference 

electrode

71 W3T161424 

O-ring

72 W3T165267 

Knurled 

nut

73 W3T161464 

Flat 

gasket

74 W3T165565 

KCl electrolyte kit, 
100ml

Item

Part No.

Designation

Содержание W3T350215

Страница 1: ...1 INSTRUCTION MANUAL DEPOLOX 5 E WALLACE TIERNAN FLOW CELL...

Страница 2: ...the express prior written con sent of Evoqua Water Technologies GmbH All information in this document is considered reliable and corresponds to the generally applicable technical standards Evo qua as...

Страница 3: ...ort and storage 18 4 3 Ambient conditions 19 4 4 Mechanical installation 19 4 5 Installation of the module 20 4 6 Removing and fitting the housing cover 23 4 7 Installing the optional strainer 23 4 8...

Страница 4: ...ng the fine filter 40 5 9 Cleaning flow rate monitor and check valve 40 5 10 Cleaning 41 6 Spare parts Accessories and Retrofit kits 42 6 1 DEPOLOX 5 E flow cell module type D01 non pressurized versio...

Страница 5: ...apters Installation Commissioning and Maintenance are intended only for trained and authorized service personnel These chap ters contain important information on the assembly configuration commissioni...

Страница 6: ...in accordance with state of the art technology and the accepted safety regulations However if the unit is used by persons who have not been adequately instructed risks to life and limb of such persons...

Страница 7: ...y inspection maintenance or repair work Then first test the disconnected compo nents to ensure they do not carry any vol tage Never use corrosive cleaning agents e g spirit scouring agents 2 4 Warrant...

Страница 8: ...temperature of the sample water The metallic sensor housing guarantees good sample water earthing The continuous hydromechanical cleaning of the electrode of the three electrode measuring cell effect...

Страница 9: ...ring module total chlorine TC O No sensor measuring module total chlorine Selection of sensor measuring module for conductivity L Sensor measuring module conductivity O No sensor measuring module cond...

Страница 10: ...ence electrode Membrane sensor for total chlorine TC3 CAN Conductivity sensor DEPOLOX 5 E 700 M Non pressurized version Part no electrode measuring cell for free chlorine chlorine dioxide or ozone pH...

Страница 11: ...uring and reference electrode 3 electrode measuring cell free chlorine X X X Membrane sensor for total chlorine TC3 CAN X X pH combined measuring and reference electrode X X X Redox combined measuring...

Страница 12: ...ater outlet in pressurized versi on also with ball valve I Sample taking unit drain J Sample water inlet K Shut off ball valve L Filter unit M Check valve housing N Flow control valve O Multi sensor P...

Страница 13: ...trode immerses in an elec trolyte solution and can also be refilled during operation The sample water that flows tangentially into the 3 electrode measuring cell ensures conti nuous hydromechanical cl...

Страница 14: ...d through a second CAN connection socket optional See design chapter 4 15 Fig 13 3 5 4 Module SiDiSens LF optional To measure the conductivity conductivity module SiDiSens LF optional is installed in...

Страница 15: ...ee drain Pressurized version max 1 5 bar Multi Sensor Switching point 21 l h 3 l h Switching hysteresis 2 l h Temperature sensor Pt1000 Sensors Measured variables Free chlorine chlorine dioxide ozone...

Страница 16: ...mbined measuring and reference electrode with platinum electrode salt reserve zirconium dioxide diaphragm polymerized solid electrolyte Ag AgCl drain system Measurement range 2000 mV Working temperatu...

Страница 17: ...ight Atmospheric pressure 75 to 106 kPa Max working height 2 000 m Relative humidity 80 non condensing Noise emission 45 dB Design Sensor electronics integrated in the flow cell for connection of the...

Страница 18: ...tion manual DEPOLOX 5 E flow cell German Englisch Instruction manual 700 M electronics module German Englisch 4 2 Transport and storage Transport The flow cell is supplied in standard packaging During...

Страница 19: ...tion site for pool water The sample water take off point must be installed in the pool return line according to standards see DIN 19643 DIN German Industrial Norm Make sure that the sample water take...

Страница 20: ...ule and flow cell with or without DIN rail Dimension drawing chapter 4 5 3 4 5 1 With DIN rail 1 Secure the DIN rail to a solid wall using the dowels and screws supplied 2 Hook the electronics module...

Страница 21: ...E flow cell 4 Installation EN 4 5 3 Dimension drawings DEPOLOX 5 E flow cell non pressurized version and electronics module example 0LWWH 0RQWDJHVFKLHQH PLGGOH PRXQWLQJ UDLO 0RQWDJHVFKLHQH PRXQWLQJ U...

Страница 22: ...22 4 Installation DEPOLOX 5 E flow cell EN DEPOLOX 5 E flow cell pressurized version and electronics module example 0LWWH 0RQWDJHVFKLHQH PLGGOH PRXQWLQJ UDLO 0RQWDJHVFKLHQH PRXQWLQJ UDLO PLQLPDO...

Страница 23: ...uld distort the measure ments Sample water inlets are classified accor ding to whether they have a hose connec tion or are fitted with rigid piping Select the sample water extraction point that guaran...

Страница 24: ...ld falsify the measure ments On the non pressurized version no back pressure is permitted in the cell body design chapter 4 8 Fig 3 The sample water outlet must be open We recommend using a funnel abo...

Страница 25: ...25 DEPOLOX 5 E flow cell 4 Installation EN 4 10 Sample water take off options Example for sample water extraction using a booster pump Only operate with sample water inlet...

Страница 26: ...26 4 Installation DEPOLOX 5 E flow cell EN Example for sample water extraction Part list on page 27 Only operate with sample water inlet...

Страница 27: ...ion 2 1 W2T505181 Screw joint 3 2 W2T505182 Screw joint 4 1 W3T171416 Strainer complete 6 1 W3T167518 Hose connection parts 7 1 W3T173160 W3T173198 Pressure gage fresh water salt water 9 1 W2T505945 B...

Страница 28: ...body 5 Push the entire electrode unit upwards 6 Refit the knurled nut on the electrolyte tank Fig 7 Cross section of cell body non pres surized version example knurled nut A Knurled nut B Felt washer...

Страница 29: ...ed version close the shut off ball valve on the sample water outlet 3 On the non pressurized version take out a plug or sensor from the cell body cover 4 On the pressurized version remove the protecti...

Страница 30: ...h the M12 cable union Fig 11 item L Connect the signal cable observing the correct colors Fig 12 and lower table Screw in the cap clock wise up to the mark Fig 11 Detail cap Fig 12 Section of 3 electr...

Страница 31: ...rom the mounting hole on the cell body cover 2 On the pressurized version unscrew the protection cap 3 Take the pH and redox combined measu ring and reference electrode out of the KCl tank with stand...

Страница 32: ...wer holding device B For the cup use the lower position C Fig 14 Rear of flow cell A Upper holding device B Position of holding device for bag C Position of holding device for cup 4 18 Commissioning A...

Страница 33: ...ew the sensors 2 Fit the pH or redox combined measuring and reference electrode into the KCl tank and stand with KCI solution 3 Store the sensors in a frost free place Fig 15 pH and redox combined mea...

Страница 34: ...To do this raise DFMe electronics module and carefully unhook it from the mounting hook 8 Open the housing of DFMe electronics module To do this release the four cover screws of DFMe electronics modu...

Страница 35: ...le 16 Refit and engage the housing cover of the flow cell 17 Switch the power supply on again Fig 18 DEPOLOX 5 E flow cell non pressurized A Mounting recess B CAN sensor cable connection C SiDiSens LF...

Страница 36: ...ard cell body cover optional The pressurized version can be retrofitted with the total chlorine membrane sensor TC3 CAN Please note that the standard cell body cover must be replaced by an optional ce...

Страница 37: ...parts kits 5 2 1 Non pressurized version 5 2 2 Pressurized version 5 3 Checking for leakage Check the flow cell including all screw connec tions for leakage every day Repair any lea kage points immed...

Страница 38: ...eaning sand electrolyte solu tion reference electrode and diaphragms 1 Switch off the power supply 2 Close the shut off ball valve D at the sample water inlet 3 On the pressurized version close the sh...

Страница 39: ...he narrow section of the KCI tank 24 Insert the plug into the electrolyte tank 25 Reinsert the electrode mount complete with electrolyte tank into the cell body The cell body s locating pin must engag...

Страница 40: ...on A Knurled nut B O ring C Fine filter 5 9 Cleaning flow rate monitor and check valve 1 Switch off the power supply 2 Drain the sample water inlet and outlet line 3 Remove the housing cover of the fl...

Страница 41: ...Abb 4 Cross section of check valve housing A Check valve housing overall B Flow cone C Spherical seat D Glass ball 5 10 Cleaning Do not use aggressive cleaning agents e g alcohol abrasive cleaners We...

Страница 42: ...Ball 31 W3T172949 O ring 32 W3T159707 Insert 30 33 W3T163739 Spherical set cpl 29 33 W3T158603 Check valve unit cpl with float cone 33 W3T172975 O ring 35 W3T158602 Filter housing 33 34 W3T166194 Fine...

Страница 43: ...ug 89 W3T161452 Felt washer transport fitting 90 W3T161453 Protection plug 300 332 DFMe electronics module 401 W3T183616 Conductivity measure ment module SiDiSens 402 W2T807965 Plastic self tapping sc...

Страница 44: ...11 18 19 16 17 81 43 45 44 40 41 39 35 9 7 8 9 51 52 53 54 80 82 36 46 47 55 10 48 56 57 58 59 60 66 69 67 70 71 72 74 75 42 12 68 15 63 64 62 61 78 79 3 406 1 405 402 89 34 33 73 Elektronikbox DFMe V...

Страница 45: ...Fine filter cpl 36 W2T505463 Plastic self tapping screw 37 W3T161450 Plug 38 W3T168859 O ring 39 W3T172041 Securing ring 40 W3T158576 Outlet drain pipe 41 W3T172997 O ring 42 W3T164597 O ring 43 W3T1...

Страница 46: ...vity measure ment module SiDiSens 402 W2T807965 Plastic self tapping screw A2 10 mm 405 W2T807968 Plastic self tapping screw 406 W3T309482 Product strip DFM 420 W2T507548 Type plate Acces W3T158743 El...

Страница 47: ...plified illustration without cable Various versions O ring slightly greased with special grease UNISILIKON L250L W2T504248 W2T504249 O Ring leicht gefettet mit Spezialfett UNISILIKON L250L W2T504248 W...

Страница 48: ...l Item 1 Item 3 Item 2 Item 4 Item 5 Item 6 Item Part No Designation 1 W3T322432 Cell body fully non pressurized version 2 W3T322435 Cell body fully pressurized version 3 W3T166171 Sample extraction u...

Страница 49: ...1 W3T166210 Electrode mount complete with electrode 2 W3T162579 Electrode unit Complete non pressurized 3 W3T159961 Plug complete non pressurized 4 W3T159824 Electrode housing non pressurized 5 W3T15...

Страница 50: ...ousing cover DFMe 303 W2T807967 Plastic self tapping screw A2 305 W3T320085 Spare PCB DFMe board DES 306 W2T504397 Plastic self tapping screw d4x10 308 W3T263401 DFMe connection M12 plug 321 W3T271603...

Страница 51: ...W3T158743 Electrode cleaning sand QK W3T158600 Measuring beaker 5 pcs Part No Designation W3T277062 LED glow stick complete DFMe LED Part No Designation W3T321512 Instruction manual 700 P electronics...

Страница 52: ...KCl bottle 250 ml W3T160410 Spare sensor cable DFMe W3T320081 W3T320082 Conductivity measurement module SiDiSens W3T183616 Sealing kit LF325 pressurized W3T166180 Membrane cap W3T365500 Plastic membra...

Страница 53: ...9 Electrolyte solution E TC 60 mS cm Bottle 1000 ml W3T171793 Measuring beaker 5 pcs W3T158600 Designation pH combined measuring and reference electrode Redox combined measuring and reference electrod...

Страница 54: ...3T172861 W3T172861 W3T169068 Locking ring W3T172891 W3T169815 W3T163436 W3T163437 Union nut W3T161502 W3T161502 W3T161502 W3T167297 Hose bushing W3T172945 W3T161501 W3T161501 W3T167293 Part No Designa...

Страница 55: ...DEPOLOX 5 E flow cell 7 EC Declaration of Conformity and Certificate 55 EN 7 EC Declaration of Conformity and Certificate...

Страница 56: ...7 EC Declaration of Conformity and Certificate DEPOLOX 5 E flow cell 56 EN...

Страница 57: ...pe Laplante Jean Philippe Laplante PRODUCTS CLASS C363106 ELECTRICAL MEASUREMENT AND TEST EQUIPMENT CLASS C363186 ELECTRICAL EQUIPMENT FOR MEASUREMENT USE Certified to US Standards Water management sy...

Страница 58: ...DEPOLOX 5 E flow cell 58 EN NOTES...

Страница 59: ...au evoqua com Canada 1 905 944 2800 wtoe can evoqua com China 86 21 5118 3777 sales cn evoqua com France 33 1 41 15 92 20 wtfra evoqua com Germany 49 8221 9040 wtger evoqua com Singapore 65 6559 2600...

Страница 60: ...Auf der Weide 10 89312 G nzburg Germany 49 8221 904 0 www evoqua com 2020 Evoqua Water Technologies GmbH Subject to modifications WT 050 810 040 DE IM 0420 W3T332382 Issue 11 0420...

Отзывы: