11 APPLICATIONS
11.1
Calibrating a temperature indicator
100.00 °C
The following procedures will be used to test and calibrate temperature indicators.
•
Switch the calibrator ON
•
Connect the instruments as in figure.
•
Set the CH2 in output mode and configure it for the desired TC or RTD type (e.g. TC type K).
To configure the channel, proceed as follows:
Press the [
SELECT
] key until “
Select Ch2
” is shown.
SELECT Ch2
Out
In
Off
Press < > to SET
Out
Thermocouple
RTD thermometer
Current
Voltage
Resistance
Press < > to SET
Thermocouple
Rji
Rje
Rjr
Press < > to SET
Rji
TcK
TcJ
TcT
TcR
TcS
Press < > to SET
15.2000
Ch1 mA
Ch2 Out TcK Rji °C
100
.
02
Change the output value by using the 6 [
] keys (You can also use the calculator mode to insert the value).