Instruction Manual MM850432 ed. 05c
62
•
Press the [
MENU
] key. The following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select “
Ramp-Memory scan
”
Ramp - Memory scan
Ramp
Memory scan
Press < > to SET
Select the “
Memory Scan
”
Memory scan
On/Off
From [1]
To [8]
Time [2]
Mode [AUTO]
Press < > to SET
Set the parameters for simulation procedure
On/Off
: runs and ends the function
Mode
: AUTO: automatic sequence using the “Time” value as interval time – MANUAL: require to press the [
HOLD
] key to go to the next step
From
: First memory to be simulated.
To:
Last memory to be simulated
Time
: interval time between 2 steps
Mode
: set the mode between the options "
Auto
" and "
Manual
"
IMPORTANT
B
EFORE RUN THE PROCEDURE
,
BE SURE TO HAVE ALL MEMORIES WITH THE SAME OUTPUT PARAMETERS
.
I
F THE STORED MEMORIES HAVE DIFFERENT PHYSICAL PARAMETERS
(
EG
.
M
V
AND M
A)
YOU
COULD SERIOUSLY DAMAGE THE LOAD
.
•
Select the “
On/Off
” key to run the procedure. The display shown the following page:
–
15.2000
Ch1 mA
Ch2 scan 0 mV
100.02
The message “
scan
” is displayed in the status box when the sequencer is running. A number shows the current step.