Instruction Manual MM850432 ed. 05
57
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select the “
Unit
” option
Unit
Temperature
Pressure [mbar]
Xscale
Press < > to SET
Select the “
XScale
” option.
XScale
X1
X2
X3
X4
Press < > to SET
Select one of the 4 available functions. (e.g. X1)
X1
Unit [°C]
Parameter [50 mA]
In Low [4.000]
In High [20.000]
Disp Low [0.0]
Press < > to SET
Select and program each of the 8 variables:
Unit
: 5 characters length can be programmed for the engineering unit
Parameter
: select the voltage or the current input range [200mV, 2V, 20V, 50mA]
In Low
: Change this value with the Zero scale
In High
: Change this value with the Input Full Scale
Disp Low
: Change this value with the Displayed low scale
Disp High
: Change this value with the Displayed High scale
Function
: Change this value with the linearization function [linear, square, sqrt, log]
Decimals
: Change this value with the number of digits to be displayed [0, 1, 2, 3, 4]
•
Press the [
ESC
] key to return in measuring mode.
5.8
Cycle & Ramp
The output channel can be programmed to simulate a ramp.
The procedure uses the Channel 2 output configuration.