Instruction Manual MM850432 ed. 05
61
5.11 Transmitter
simulator
The calibrator can be programmed for simulate a signal or a pressure transmitter. The input signal is transformed to a 0-20mA or a 4-20mA output signal.
The procedure uses channel configuration. Before run the procedure the source channel has to be set.
•
Press the [
MENU
] key. The following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select the “
Advanced
” option
Advanced
Switch Test
Leak Test
HART
Trx Simulator
Graph
Press < > to SET
Select the “
Trx Simulator
” option
Trx simulator
On/Off
Source [Ch1]
In values
Out values
Function [linear]
Press < > to SET
Press the [ ] key to display the other variables
Trx
In value
Out value
Function [linear]
In error [Error]
Sampling time [1]
Press < > to SET
On/Off
: RUN and STOP the function
Source
: Input channel (Ch1, P1, P2, PE, PA, RH%, T)
In value
: Allows to program the low and full scale for the Trx source
Out value
: Allows to program the low and full scale for the mA output (max. range 0 to 50mA)
Function
: Linearization mode (Linear, square, sqrt, log)
In error
: Error message for input out of range (out=L/H, error)
Sampling time
: It is possible to choose the interval between input data acquisition (0.1, 1 or 10 seconds)
5.12 Memory
Scan
The output channel (CH2) can be programmed for simulating a sequence of 10 values. This procedure uses the Channel 2 as output channel and the 1-9 memories.
Before run this feature, you have to set output values using
consecutive memories