IN Thermocouples
TC TYPE RANGE
RESOL. ACCURACY
Plus model
Tc J
-210 to 1200 °C
-35 to 2200 °F
0.01 °C
0.01 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc K
-270 to 1370 °C
-450 to 2500 °F
0.01 °C
0.01 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc T
-270 to 400 °C
-450 to 760 °F
0.01 °C
0.01 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc R
-50 to 1760 °C
-60 to 3200 °F
0.1 °C
0.1 °F
±(0.01% +0.2°C)
±(0.01% +0.4°F)
Tc S
-50 to 1760 °C
-60 to 3200 °F
0.1 °C
0.1 °F
±(0.01% +0.2°C)
±(0.01% +0.4°F)
Tc B
50 to 1820 °C
140 to 3300 °F
0.1 °C
0.1 °F
±(0.01% +0.3°C)
±(0.01% +0.6°F)
Tc C
0 to 2300 °C
32 to 4150 °F
0.1 °C
0.1 °F
±(0.01% +0.2°C)
±(0.01% +0.4°F)
Tc G
0 to 2300 °C
32 to 4150 °F
0.1 °C
0.1 °F
±(0.01% +0.3°C)
±(0.01% +0.6°F)
Tc D
0 to 2300 °C
32 to 4150 °F
0.1 °C
0.1 °F
±(0.01% +0.3°C)
±(0.01% +0.6°F)
Tc U
-200 to 400 °C
-330 to 760 °F
0.1 °C
0.1 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc L
-200 to 760 °C
-330 to 1400 °F
0.1 °C
0.1 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc N
-270 to 1300 °C
-450 to 2380 °F
0.1 °C
0.1 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc E
-270 to 1000 °C
-450 to 1840 °F
0.1 °C
0.1 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Tc F
0 to 1400 °C
32 to 2560 °F
0.1 °C
0.1 °F
±(0.01% +0.1°C)
±(0.01% +0.2°F)
Note: 0.1°C resolution with temperature lower than –200°C
Selection °C/°F/K: through the configuration procedure
Resolution: 0.01°C / 0.01°F
Temperature scale: ITS90 and IPTS68 selectable
Reference junction compensation:
internal automatic from -10 °C to +55 °C
external adjustable from -50 °C to +100 °C
remote with external Pt100 from -10°C to +100 °C (only on XP model)
Rj compensation drift: ± 0.002°C/°C (from -10 °C to +45 °C)
Rj accuracy: ±0.05°C @ 25°C / ±0.005 °C/°C
Input impedance (Tc ranges): >10 M
Ω
Environmental condition module (option)
Temperature:
Sensor type: Pt100
Range:
Accuracy:
Relative Humidity:
Sensor type:
Range: 0 to 95% RH
Accuracy: ±2% RH
Pressure:
Sensor type:
Range:
Accuracy: ±2 mbar
Math functions
Calculation functions: hold, max, min, offset, zero, average
In/Out data memory: 10 data with manual or automatic recall
Convert function: displays the electrical equivalent of the engineering unit
Scale factor: setting with zero and span programmable
Square root: in combination with scale factor
DataLogger
Sources: Voltage, Current, Temperature, Resistance
Sampling time: 1 sec
Memory: >1500 readings complete with date and time
General
Accuracy: the above accuracies are stated for 365 days and includes non-linearity,
hysteresis, and repeatability. The average temperature coefficient, inside the
temperature compensated range is ±0.002 of rdg/°C (w.t.r. +23°C / +73°F).
Compensation temperature range: 0 to 45°C (+32 to +113°F)
Calibration: self learning technique with automatic procedure
Channel 1-Channel 2 insulation: 250 Vdc
Common mode rejection: 140 dB at ac operation
Normal mode rejection: 60 dB at 50/60 Hz
Temperature stability:
for temperature exceeding the band +18°C to 28°C
Span: ±8 ppm/°C
Zero: ±0.2
µ
V/°C
Measurement sampling time: 250 ms
Display: graphic LCD display with automatic and manual backlight device
Digital interface: full bi-directional RS232
Power supply: external charger and rechargeable Ni-MH battery
Battery life (typical):
10 h on Tc and mV input/output (backlight Off)
4 h with 20 mA simulation (backlight Off)
Recharging time (typical): 5 h at 90% and 6 h at 99% with instrument switched off.