Instruction Manual MM850432 ed. 05c
76
8.1.1
Data logger: data storing
This section of
Utilities Manager
allows downloading on PC data stored in the internal memory of your MicroCal.
First of all, you should have a data file realized with the instrument itself (Data Logging)
The calibrator can be programmed for memory store input channels measurements. Each measure is recorded complete of the current data and time.
The procedure uses channels configuration. Before run the procedure the channels to be logged have to be set.
•
Press the [
MENU
] key. The following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select the “
Data logger
” option
Data Logger
On/Off
Channels
Time
Name [MyLog]
Replay
Press < > to SET
On/Off
: RUN and STOP the logging
Source
: Select the channels for recording. Available settings are: T+RH%+P, (only if optional environmental module is installed),
PL+PH+Pe (if internal and/or external pressure sensor are installed), CH1+CH2.
Time
: Data Acquisition interval (from 1 second to 60 minutes).
Name
: File name. Use keypad to input the file name.
Replay
: Replay function allows recalling stored data in a log file (see next chapter 8.1.2)
•
Select the “
On/Off
” key to run the procedure. The display shows the following page:
15.2000
Ch1 mA
Ch2 Out TcK Rji °C
100.02
L
The symbol “
L
” on the message box is displayed if data logging is running.
•
Return back to menu and in Data Logger item select newly “
On/Off
” to stop the procedure.
Source
Ch1
Ch2
Pl
Ph
Pe
Press < > to SET