Instruction Manual MM850432 ed. 05
53
NOTE
A
BOUT THE GENERATION OF FREQUENCY
/
PULSE
,
SEE PAR
.
4.6.2
5.6
Source mode: signals generation
The calibrator can generate (source) all the electrical signals as shown in technical specifications paragraph.
•
Press the [
SELECT
] key until “
SELECT CH2
” is shown.
SELECT Ch2
Out
In
Off
Press < > to SET
Select the “
OUT
” option. The display shows the following page
OUT
Thermocouple
RTD thermometer
Current
Voltage
Resistance
Press < > to SET
NOTE
P
ROCEED AS DESCRIBED IN THE MEASURE PAR
.
4.5
FOR THE SELECTION OF VARIETIES OF SIGNAL GENERATIONS
To insert the numeric value for a generated signal, it is enough to insert it directly using [ ], [ ] (numeric pad). Each key is located immediately below the digit it acts on: a single click
of [ ] increases the digit of 1 unit, vice versa a single click of [ ] decreases this digit of 1 unit.
On the other hand, it is possible to insert a value with calculator function:
•
Press
[
CALC
], the calculator box and the related symbol appear
•
Insert a value with the numeric pad
•
Confirm by pressing [
MENU
], the value is then transcribed on the output channel
•
Press
[
ESC
] to exit from calculator function
1 0 .00
0.000
0.000000