Instruction Manual MM850432 ed. 05
59
See the following explanatory figure:
•
Select the “
On/Off
” key to run the procedure. Press [
ESC
]: the display shows a page like that:
15.2000
Ch1 mA
Ch2 Ramp mV
100.02
The “
ramp
” massage is displayed on the channel 2 while the ramp is running.
The phases of the ramp are signed by the arrows beside the “ramp” message: = ascent, = soak up, = slope down,
= soak down
•
Press the [
HOLD
] key to pause and to continue the procedure.
Example
:
Ch2 on V
From=0, to=1, step=0.1, t up=5, t soak up=3, t down=4, t soak down=4
The simulation starts from 0V, goes to 1V in 5sec with 0.1V steps. After the ascent, the signal is kept at 1V for 3sec and goes down to 0V in 4sec. After other 2sec, the cycle is redone
for a number of times equal to “n.cycles”.