Instruction Manual MM850432 ed. 05c
66
5.15 Alarms
It is possible to enable or disable alarm’s thresholds initially set up by the firm (system alarms) or set custom alarm definable by the user.
•
Press
[
MENU
], the following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select “
Configuration
”
Configuration
Display
Power On [Set Ch]
Date/time
Rejection [50 Hz]
Baud rate [9600]
Press < > to SET
Scroll down and select "
Alarm
"
Configuration
Date/time
Rejection [50 Hz]
Baud rate [9600]
OC Test [Enable]
Alarm
Press < > to SET
Alarm
General [Enable]
User
Press < > to SET
•
Select
General
and then
Enable
or
Disable
the general system alarms
NOTE:
WHEN YOU DISABLE THE ALARMS
,
AT NEXT TURNING ON THE SELECTION WILL BE LOSE AND ALARMS BECOME ACTIVE
.
See par.10.1 about error messages.
•
Select
User
to set alarms definable by the user:
User
OnOff [off]
Channel [ch1]
Min [0]
Max [100]
Press < > to SET