Instruction Manual MM850432 ed. 05
91
Result
Found
Left
Press [ ] to SET
•
Now it is possible to run the procedure by pressing “
Run procedure
” and to wait for its completion:
OFF_VtoV 1/1
Ref:5.0000V+0.0125
Ref:Out +5.0000 V
Act:In +5.0002 V
Limit: +0.0035
Error: +0.0002
Procedure complete
Press [ENTER]
When the Reference and/or the Actual values are out of limits, the correspondent values will be in reverse mode.
Eventually, abort the procedure by pressing [
ESC
] key. A confirmation message will be shown on display.
•
Finally, press [
ENTER
] as required at the end of the procedure. The calibrator returns to the measuring mode.
•
Returning to procedure menu by pressing [
CAL PROC
] and selecting “
Display results
”, it is possible to check step by step the results of the calibration (press [
ENTER
] to change
the results pages).
SELECT Procedure
Tag [Tag1]
Results [Found]
Display procedure
Display results
Run procedure
Press [ ] to SET