Instruction Manual MM850432 ed. 05c
56
If you select “
Frequency
” as output, the display shows:
Frequency
Square
Sinusoidal
Triangular
Amplitude [5]
Sin offset [0]
Press < > to SET
•
Select
the
Amplitude
•
Select the offset with “
Sin offset
”
•
Select the desired waveform: square, sinusoidal, triangular
•
Select the range for the frequency :
Sinusoidal
200Hz
2kHz
20kHz
Press < > to SET
•
Press the [ ] key to select the option and to enter measure mode
If you select “
Pulse
” as input, the display shows:
Pulse
OnOff
Time
Mode [One shot]
Amplitude [5]
Press < > to SET
•
Define the time (minutes and seconds), over that the generation is performed
with
Time
•
Select the mode for pulse generation: “
One shot
” or “
Continuous
”, depending
on which kind of generation you want: over a single time range or continuously
•
Select the
Amplitude
•
Select
OnOff.
The instrument comes back automatically at the measuring
page and channel 2 expresses the value in p/t. Hence it is possible to directly
set the number of pulses in time unit.
When the instrument is in the measures page, it appears in Hold mode. Unlocking
hold by pressing [
HOLD
], the generation starts with those previously set
parameters.
5.7 X-Scaling
setting
The calibrator can measure voltage or current and simply display it or convert it using the “XScale” function and then display the measured value via a conversion factor.
•
Press the [
MENU
] key. The following page is shown: