Instruction Manual MM850432 ed. 05
43
IMPORTANT
P
LEASE SELECT
4
WIRE CONNECTION IF YOU HAVE A
2
WIRE
RTD
AND USE THE ELECTRICAL CONNECTION AS SHOWN ON
2.4
CHAPTER
.
•
Press the [ ] key to confirm the selection. Press the [
ESC
] key to return in measure mode without changes.
3 wire
Pt100 IEC
Pt200 IEC
Pt500 IEC
Pt1000 IEC
Pt100 OIML
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the requested type of RTD.
•
Press the [ ] key to confirm the selection and going in measuring mode. Press the [
ESC
] key to return in measure mode without changes.
5.5.4
Temperature Scale Setting
•
Press the [
MENU
] key. The following page is shown:
MENU
Configuration
Unit
Ramp-Memory scan
Data logger
Advanced
Press < > to SET
Select the “
Unit
” option
Unit
Temperature
Pressure [mbar]
Xscale
Press < > to SET
Select the “
Temperature
” option
Temperature
Unit [°C]
Rj Ch1 ext [0.0]
Rj Ch1 ext [0.0]
Scale [ITS 90]
Press < > to SET
Select the “
Scale
” option
Temperature
IPTS 68
ITS 90
Press < > to SET
•
Press the [ ] and [ ] keys to highlight the temperature scale used (ITS90 or IPTS68).
•
Press the [ ] key to confirm the selection and going in measuring mode. Press the [
ESC
] key to return in measure mode without changes.