Device Profile Area
Manual • Doc.-No.: C.3032.21 / Rev. 1.1
CAN-CBX_Pt100
Page 88 of 120
8.10.7 AI Physical Unit (6131
h
)
Index
Sub-
index
Description
Value range
Default
Data type
Access
mode
6131
h
0
Number_of_entries
4
4
unsigned 8
ro
1
AI_physical_unit_1
28 0000
h
,
2D 0000
h
unit
SensorType
_1 unsigned 32
ro
2
AI_physical_unit_2
28 0000
h
,
2D 0000
h
unit
SensorType
_2 unsigned 32
ro
3
AI_physical_unit_3
28 0000
h
,
2D 0000
h
unit
SensorType
_3 unsigned 32
ro
4
AI_physical_unit_4
28 0000
h
,
2D 0000
h
unit
SensorType
_4 unsigned 32
ro
The default values unit
SensorType
_x (x = 1 - 4) depend on the sensor-type set.: With the default-setting of object 6110
h
=10 000 (2710
h
= resistance measurement) it would be 2 621 440
d
= resistance in Ohm.
Description of the variable AI_physical_unit_x (x = 1...4):
These variables contain the physical units and prefixes for the measuring channels used. The value
of the variables is determined by the sensor type specified in object “AI_sensor_type” (object
6110
h
).
The structure of the variable is defined in CiA DS 404. The coding of the physical units and prefixes
is done according to CiA303-2 [5].
The structure is described on the following page.