Manufacturer Specific Profile Area
CAN-CBX_Pt100
Manual • Doc.-No.: C.3032.21 / Rev. 1.1
Page 111 of 120
8.11.10.3 Gain Correction at I
LOW
and I
HIGH
(2502
h
, 2503
h
)
Index
Sub-
index
Description
Value range
Default
Data type
Access
mode
2502
h
0
Number_of_entries
4
4
unsigned 8
ro
1
gain_correction_Ilow_1
8000
h
...7FFF
h
*
6)
integer 16
rw
2
gain_correction_Ilow_2
8000
h
...7FFF
h
*
6)
integer 16
rw
3
gain_correction_Ilow_3
8000
h
...7FFF
h
*
6)
integer 16
rw
4
gain_correction_Ilow_4
8000
h
...7FFF
h
*
6)
integer 16
rw
2503
h
0
Number_of_entries
4
4
unsigned 8
ro
1
gain_correction_Ihigh_1
8000
h
...7FFF
h
*
6)
integer 16
rw
2
gain_correction_Ihigh_2
8000
h
...7FFF
h
*
6)
integer 16
rw
3
gain_correction_Ihigh_3
8000
h
...7FFF
h
*
6)
integer 16
rw
4
gain_correction_Ihigh_4
8000
h
...7FFF
h
*
6)
integer 16
rw
*
6)
The default values have been determined individually for every module at the calibration. Thus these values are
not described in this general documentation.
Description of the parameter gain_correction_Ilow_x and gain_correction_Ihigh_x (x = 1...4):
These objects contain the gain-correction factors, that are determined during calibration of the
individual channels for I
low
and I
high
at PGA=1.
Value range: 8000
h
... 7FFF
h