Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
83
37.16 Internal 0.2µ filter for N
2
gas
The round-shape in line 0.2µ HEPA filter for N
2
gas removes any particles found in the
incoming N
2
gas. Failure to use the HEPA filter may cause damage to the high precision
flow sensor, calculate the amount of N
2
gas entering the system, and disturb the N
2
regu-
lator system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming N
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
37.17 Firmware update
If Esco Medical has released a newer version of the firmware, this should be installed on
the Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators during the yearly
scheduled service.
Please refer to the service manual for instructions on how to update the firmware.
38 The Installation guide
This section describes when and how to install the Mini MIRI
®
Dry and Mini MIRI
®
Hu-
midity multiroom IVF incubator in the IVF clinic.
38.1 Responsibilities
All technicians or embryologists installing the Mini MIRI
®
Dry or Mini MIRI
®
Humidity
multiroom IVF incubator must identify problems and perform any necessary calibrations,
adjustments and maintenance.
Installation personnel performing MEA (Mouse Embryo Assay) must be thoroughly fa-
miliar with the MEA and all functions of the instrument, calibration and testing proce-
dures, and instruments used in the instrument's testing. MEA test is a supplemental in-
stallation test and is not mandatory.