Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
20
3.
Connect the mains cable to the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom
IVF incubator.
4.
Connect gas lines.
5.
Set the gas pressure on the external gas regulator at 0.4
–
0.6 bar (5.80
–
8.70 PSI).
6.
Switch on the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator in
the back.
7.
Observe for standard functionality.
8.
Let the unit warm up and stabilize for 20 min.
9.
Follow the guidelines in the Validation guide (see section
“33
The validation
guide”)
10.
Complete user training and finish reading instructions.
11.
After a burn-in phase of 24 hours, the unit is ready for use IF the testing is success-
ful.
Clean and disinfect the device before use. It is not delivered sterile or in a clin-
ically acceptable cleanliness state. Consult the cleaning instructions section in this
manual for the manufacturer's recommended guidelines!
10 Mains connection
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators come with a
detachable mains power cord. The power cord is prepared for the country in which the
unit is intended to be used.
The ON/OFF switch provides the user with a means to isolate the Mini MIRI
®
Dry and
Mini MIRI
®
Humidity multiroom IVF incubator from the mains.
Do not defeat the safety purpose of the grounding-type plug! A grounding-type
plug has two blades and a prong, which is provided for your safety. If the provided
plug does not fit into your outlet, consult an electrician to replace the outlet.
The power requirement is 230V 50 Hz OR 115V 60Hz. The built-in power supply has a
switch mode that automatically adjusts to the correct mains power between 100V-240V
AC 50-60 Hz.
Figure 10.1
Power supply