![Esco Medical Mini MIRI Dry Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/esco-medical/mini-miri-dry/mini-miri-dry_user-manual_2426159048.webp)
Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
48
Place the dish where it fits the pattern. The heat optimization plates can be applicable for
Nunc
®
, Falcon
®
, Sparmed
®
and VitroLife
®
dishes.
Use only the correct type of heat optimization plates for your dishes.
Figure 22.1
Heat optimization plate
Never incubate without the plates in place and
never use non-Esco Medical
heating optimization plates. It may cause dangerous and unpredictable tempera-
ture conditions that may be harmful to the specimens.
23 Humidification
23.1 Mini MIRI
®
Dry multiroom IVF incubator
The Mini MIRI
®
Dry multiroom IVF incubator is used for culture with mineral oil or Par-
affin overlay. Set the culture mode for "Oil culture". It cannot be used for open culture
mode.
The Mini MIRI
®
Dry multiroom IVF incubator must not be irrigated. Humidification of the
Mini MIRI
®
Dry multiroom IVF incubator will damage the device
–
condensation will
block internal pipes and damage electronic parts.
Open culture is possible in a 4-well (or similar type of dish) in volumes not under 0.8 mL
per well without oil overlay for up to a maximum of 4 hours. The Osmolality will change
rapidly after that and reach over 300 mOsm/kg. In a more extended period of time risk
of osmolality changes in media will increase rapidly.
In "Oil culture" mode, the lid temperature is kept 0.2
–
0.3 °C above the temperature set-
point. In "Open culture" mode, the lid temperature will be increased by an additional 1.0
°C above the temperature setpoint (thus making the lid 1.2
–
1.3 °C warmer than the in-
sert surface).