It denotes that the sensor in compartment 2 has failed. As a safety precaution, the heating
of the affected area will be switched off.
15.2 Gas level alarms
15.2.1 CO
2
alarms
The CO
2
gas level alarm is activated if the concentration of the CO
2
gas deviates more than
± 1% from the setpoint.
Remember that changing the setpoint more than ± 1% from the current gas
level will result in a gas level alarm. The same goes for all calibration adjustments.
CO
2
gas % is too low:
CO
2
gas % is too high:
The display will lock on the alarm condition and will stop alternating between the stand-
ard status messages. If the mute key is pressed, the display will shift to normal status and
show the parameters for 5 minutes until the audio alarm comes back on again. The mute
alarm key will still show the alarm condition by blinking red while the alarm is muted.
15.2.2 O
2
alarms
The O
2
gas level alarm is activated if the concentration of the O
2
gas deviates more than
±1% from the setpoint.
Remember that changing the setpoint more than
±
1% from the current gas
level will result in a gas level alarm. The same goes for all calibration adjustments.
O
2
% is too low: