11 Gas connections
There are two gas inlets on the back of the unit. These ports are marked "CO
2
100% Inlet"
and "N
2
100% Inlet".
Figure 11.1
Gas inlets on the back of Mini MIRI
®
Dry and
MIRI
®
Humidity multiroom IVF incubators
CO
2
inlet should be connected to a 100% concentration of CO
2
. CO
2
control in the com-
partment is available in the range from 1.9% to 9.9%.
The N
2
inlet should be connected to 100% concentration N
2
if low oxygen conditions are
required. The O
2
control in the compartments is available in the range from 3.9% - 19.9%
by infusing N
2
.
The premixed gas inlet should be connected to the CO
2
inlet.
The inlet's gas pressure should be between 0.4
–
0.6 bar (5.80
–
8.70 PSI) and
it must be kept stable!
Always use a high-quality pressure regulator that can be set with the required precision
for both gases.
Figure 11.2
Pressure regulator
Connect the CO
2
gas to the CO
2
inlet with a suitable silicone tube. Ensure that the tube is
fastened with a clip so that it does not accidentally loosen itself during sudden pressure
fluctuation. Use the supplied 0.2µ HEPA filter on the gas line just before the inlet on the
Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubator. Notice the direction.