![Esco Medical Mini MIRI Dry Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/esco-medical/mini-miri-dry/mini-miri-dry_user-manual_2426159050.webp)
Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
50
The compartment's temperature conditions can be continuously logged through the ex-
ternal connectors on the unit's side without compromising its performance.
Any logging system that uses standard PT-1000 sensors may be used.
Esco Medical can supply an external logging system (MIRI
®
–
GA12) for the sensors.
25 Gas level validation
Gas concentration in each compartment of the Mini MIRI
®
Dry and Mini MIRI
®
Humidity
multiroom IVF incubator can be validated by taking a gas sample from one of the 2 gas
sample ports on the unit's side, using a suitable gas analyzer.
Figure 25.1
Gas sample ports
Each sample port is directly connected to the corresponding compartment with the same
number. The gas sample will be taken ONLY from the specific compartment.
An external automatic gas sampler can be connected to the ports for continual
validation.
Before any gas measurement, make sure that the lids had not been opened for
at least 5 minutes.
Taking out a large sample volume may affect gas regulation.
Make sure that the gas analyzer is calibrated before use.
26 Alarm switch for an external system
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubator can be connected
to an external monitoring system, ensuring maximum safety, especially during nights and
weekends. The Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator is