Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
59
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Performance
A
The power-frequency magnetic
fields' level should be charac-
teristic of a specific location in a
commercial or hospital envi-
ronment.
Guidance and manufacturer's declaration
–
electromagnetic immunity
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators are intended for use in the elec-
tromagnetic environment specified below. The customer or the user of the Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubator should ensure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment-
guidance
Conducted
RF IEC
61000-4-6
Radiated
RF IEC
61000-4-3
10 Vrms
150kHz to
80 MHz in
ISM bands
3 V/m
80 MHz to
2.5 GHz
3V/m from
80MHz to
2.5 GHz
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of Mini MIRI
®
Dry
and Mini MIRI
®
Humidity multiroom
IVF incubators, including cables, than
the recommended separation dis-
tance calculated according to the
equation, applicable to the transmit-
ter’s
frequency.
Recommended separation dis-
tance
d
= 0.35
P
d
= 0.35 P, 80MHz to 800MHz
d
= 0.7 P, 800MHz to 2.5GHz
P is the maximum power output rat-
ing of the transmitter in watts (W)
according to the transmitter manu-
facturer,
d
is the recommended sep-
aration distance in meters (m).
As determined by an electromag-
netic site survey, field strengths from
fixed RF transmitters should be
lower than the compliance level in
each frequency range.
Interference may occur in the vicin-
ity of the equipment.