Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
75
These components must be replaced in the time intervals specified below. Failure to fol-
low these instructions may, in the worst-case scenario, result in damage to the specimens
in the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator.
Warranty void if service intervals are not followed according to table 37.1.
Warranty void if non-original parts are used or non-trained and non-author-
ized personnel carry out the servicing.
The table below shows time intervals in which components must be replaced:
Table 37.1
Service interval plan
Component name
Every
month
Every 3
month
Every
year
Every 2
years
Every 3
years
Every 4
years
VOC/HEPA filter capsule
5
X
Humidification Bottle
6
X
HEPA in-line filter for CO
2
gas
X
HEPA in-line filter for N
2
gas
X
O
2
sensor
X
CO
2
sensor
X
UV light
5
X
Cooling fan
X
Internal gas pump
5
X
Pump module
6
X
Proportional valves
X
Gas lines
X
Flow sensors
X
Pressure regulators
X
Internal 0.2µ filter for CO
2
X
Internal 0.2µ filter for N
2
X
A firmware update (if a new version
has been released)
X
37.1 VOC/HEPA filter capsule (applicable only for the Mini MIRI
®
model)
The VOC/HEPA filter capsule is placed on the Mini MIRI
®
Dry multiroom IVF incubator
unit's back for easy replacement. In addition to the active carbon component, this capsule
also has an integrated HEPA filter inside, enabling it to remove particles and volatile or-
ganic compounds from the air being re-circulated to the compartments. Because of the
carbon component's lifespan, all VOC filters' lifetime is limited, and they must be replaced
5
Only for the Mini MIRI
®
Dry model
6
Only for the Mini MIRI
®
Humidity model