![Esco Medical Mini MIRI Dry Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/esco-medical/mini-miri-dry/mini-miri-dry_user-manual_2426159027.webp)
Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
27
Rotate the rotatory button (
↻
) to scroll further down in the menu.
Press the rotatory button to enter the Service sub-menu.
13.4.1 Temperature sub-menu
Press the (
⇨
) key on the temperature menu to enter the temperature sub-menu.
Calibrate holding down the SP key and using (
⇧
) and (
⇩
) keys to adjust.
Move to the next sub-menu item with (
⇩
) key or one step up with (
⇧
) key.
Each compartment has two internal temperature sensors. One in the compart-
ment lid and another in the compartment bottom.
Example
–
how to calibrate the temperature:
The temperature has to be measured with a suitable and calibrated device. With a quality
thermometer, it has been estimated that T1 is 37.4 °C. Locate "T1 CAL" in the sub-menu
and press and hold the SP key. The display should show:
Adjust the temperature by pressing the (
⇧
) key 4 times while still holding the SP key
down. The display will show the steps from 37.1, 37.2, 37.3 and 37.4. When temperature
equals the measured temperature, let go of the SP key. The new value is stored and the
temperature sensor for the T1 area has been modified.
Calibration procedure is the same for T1
–
T4.