3 Indication for use
The Esco Medical MIRI
®
family`s multiroom IVF incubators are intended to be used to
pro-vide a stable culture environment at or near body temperature and CO
2
/N
2
or pre-
mixed gases and humidification for the development of gametes and embryos during in
vitro fertilization (IVF) / assisted reproduction technology (ART) treatments.
4 About the product
The Esco Medical Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators
are CO
2
/O
2
incubators.
Direct warming of the dishes in the chambers gives superior temperature conditions in
comparison to conventional multiroom IVF incubators.
The temperature in the compartment will remain stable up to 1 °C (even when a lid is
open for 30s) and will recover within 1 min after the lid is closed.
The Esco Medical Mini MIRI
®
multiroom IVF incubators have 2 completely separate cul-
ture heat chambers. Each chamber has its own heated lid and warming plate for Petri
dish. Mini MIRI
®
Dry and Mini MIRI
®
Humidity capacity for 35 mm Petri dish is 16 pcs.
And for 60 mm and 4-well Petri dishes
–
8 pcs.
To ensure maximum performance, the system of Mini MIRI
®
Dry and Mini MIRI
®
Humid-
ity multiroom IVF incubators have 4 separate PID temperature controllers. They control
and regulate temperature in culture chambers and lids. Compartments do not affect each
other's temperatures in any way. The top and the bottom of each compartment is sepa-
rated with a PET layer so that the lid temperature would not affect the bottom. For vali-
dation purposes, each compartment has a PT-1000 sensor built in. The circuitry is sepa-
rated from the unit's electronics so it remains a truly separate validation system.
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubator needs 100% CO
2
and 100% N
2
or premixed gas (for instance 5% CO
2
; 5% O
2
and 90% N
2
) in order to be
able to control the CO
2
and O
2
concentrations in the culture chambers.
The multiroom IVF incubator needs 100% CO
2
and 100% N
2
or premixed gas (for in-
stance, 5% CO
2
; 5% O
2
and 90% N
2
) in order to be able to control CO
2
and O
2
concentra-
tions in the culture chambers.
A dual beam infrared CO
2
sensor with extremely low drift rates controls the CO
2
level. A
chemical medical grade oxygen sensor controls the level of O
2
.