Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
70
If the device will not be clinically operational with the O
2
regulation activated, but there
is N
2
gas available, the test should be conducted with O
2
regulation switched on and with
N
2
gas supply.
If the N
2
is not available, the test can be done without it.
Make sure that the Esco Medical data logger software is running.
Check that parameter are logged and give a meaningful reading. Let the device run with-
out interfering for at least 6 hours. Analyze the results on the graphs.
Pass I:
Internal sensor temperature variation from set point is within ± 0.1 °C absolute.
Pass II
Internal sensor CO
2
concentration variation from setpoint within ± 0.2% absolute.
Pass III:
Internal sensor N
2
concentration variation from setpoint within ± 0.2% absolute.
Pass IV:
Gas flow CO
2
less than 2 l/h (for Mini MIRI
®
Humidity model
–
less than 4 l/h).
Pass V:
Gas flow N
2
less than 8 l/h (for Mini MIRI
®
Humidity model
–
less than 12 l/h).
35.11 Cleaning
Always validate the cleaning procedures locally or consult the manufacturer
or the distributor for more guidance.
After the testing has been conducted successfully, it should be cleaned again before the
device is taken into clinical use.
Inspect the unit for physical signs of dirt or dust. The unit should look generally tidy.
•
Clean the unit externally with a lint-free cloth moistened with a 70% alcohol solu-
tion.
•
Switch off the device and remove the mains lead.
•
Remove all heat optimization plates and clean them with a lint-free cloth soaked
with a 70% alcohol solution.
•
Wipe the inside of the 2 compartments with a lint-free cloth moistened with a 70%
alcohol solution.
•
Wipe the lids in the same way.
•
Let the lids remain open for 5 minutes.
•
Wipe the 2 compartments and the heat optimization plates with a lint-free cloth
soaked in sterilized water.
•
Wipe the lids in the same way.
•
Let the lids remain open for 10 minutes.
•
Re-insert the heat optimization plates.
•
Close the lids.