![Esco Medical Mini MIRI Dry Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/esco-medical/mini-miri-dry/mini-miri-dry_user-manual_2426159019.webp)
Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
19
12.
Place this unit in a location with adequate ventilation to prevent internal heat
build-up. Leave at least 10 cm clearance from the rear, 30 cm from the top and 20
cm from left and right to prevent overheating and allow access to the ON/OFF
switch in the back.
13.
This unit is intended for indoor purposes only.
14.
The unit must be connected to a suitable uninterrupted power supply (UPS)
source.
8.3 Post installation
1.
Refer all servicing procedures to qualified service personnel.
2.
Servicing is required according to the service manual as well as cases when the
device has been damaged in any way, e. g. suppose the apparatus has been
dropped, exposed to rain or moisture or does not operate normally. The Mini
MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators contain high volt-
age components that may be hazardous.
3.
Unplug this device during lightning storms or when unused for an extended pe-
riod of time.
4.
Protect the power cord from being walked on or pinched, particularly at the plug,
convenience receptacles and the point where it exits from the apparatus.
5.
Perform temperature and gas calibration at the intervals described in the manu-
als.
6.
Never leave the lids open for more than 10 sec while in use.
7.
VOC/HEPA filters must be changed every 3 months (do not apply for Mini MIRI
®
Humidity multiroom IVF incubator).
8.
A maintenance plan must be fulfilled to keep the device safe.
9.
NEVER block gas supply holes in the compartment.
10.
Ensure that CO
2
and N
2
gas supply pressures are kept stable at 0.4
–
0.6 bar (5.80
–
8.70 PSI).
11.
Never use any other except Esco Medical filter. Otherwise, the warranty will be
void.
12.
Do not use the device without a proper Esco Medical VOC/HEPA filter attached
(do not apply for Mini MIRI
®
Humidity multiroom IVF incubator).
9 Getting started
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity
multiroom IVF incubators must
be installed by authorized and trained personnel only!
1.
Follow the guidelines in the safety instructions and warnings section.
2.
Connect the mains cable to the UPS.