Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
54
If multiple temperature alarms go off:
•
Remove the samples from the affected compartments. They can be relocated to
any of the other compartments, which happens to be unoccupied. All compart-
ments are separate so that the remaining ones will function normally.
•
Alternatively, remove the samples from all the affected compartments and place
them in an alternative or backup device that is not affected by the problem.
If the CO
2
level alarm goes off:
There will be a 30-minute-long interval during which the user can assess if the condition
is temporary or permanent. If the state is permanent, remove all the samples and place
them in an alternative or backup device that is not affected by the problem. If the condi-
tion is temporary and the CO
2
level is low, keep the lids shut. If the state is temporary and
the CO
2
level is high, open a few lids to vent out some CO
2
.
If the O
2
level alarm goes off:
Usually, no emergency procedures are necessary in this case. If the condition is judged to
be permanent, it may be advantageous to switch off O
2
regulation in the menu.
If the CO
2
pressure alarm goes off:
Inspect the external gas supply and gas supply lines. If the problem is external and not
readily fixed, follow the guidelines under the section "CO
2
pressure alarm".
If the O
2
pressure alarm goes off:
Inspect the external gas supply and gas supply lines. If the problem is external and not
readily fixed, follow the guidelines under the "O
2
pressure alarm" section.
In case of a gas
pressure alarm on the unit:
Inspect the external gas supply and the gas supply lines. If the problem is external and
not readily fixed, or if the problem is internal, follow the guidelines under gas level
alarm.