Operation options
Proline Promass F 300 EtherNet/IP
68
Device
WLAN interface
IP address
If the IP address of the device is not known:
• The IP address can be read out via local operation:
Diagnostics → Device information → IP address
• Communication with the Web server can be established via the default IP
address 192.168.1.212.
The DHCP function is enabled in the device at the factory, i.e. the device
expects an IP address to be assigned by the network. This function can be
disabled and the device can be set to the default IP address 192.168.1.212:
set DIP switch No. 4 from
OFF
→
ON
.
Set the default IP address → 48.
8.4.3
Establishing a connection
Via service interface (CDI-RJ45)
Preparing the measuring device
1. Depending on the housing version:
Release the securing clamp or securing screw of the housing cover.
2. Depending on the housing version:
Unscrew or open the housing cover.
3. The location of the connection socket depends on the measuring device and the
communication protocol:
Connect the computer to the RJ45 connector via the standard Ethernet connecting
cable .
Configuring the Internet protocol of the computer
The IP address can be assigned to the measuring device in a variety of ways:
• Dynamic Host Configuration Protocol (DHCP), factory setting:
The IP address is automatically assigned to the measuring device by the automation
system (DHCP server).
• Hardware addressing:
The IP address is set via DIP switches → 47.
• Software addressing:
The IP address is entered via the
IP address
parameter (→ 104) .
• DIP switch for "Default IP address":
To establish the network connection via the service interface (CDI-RJ45): the fixed IP
address 192.168.1.212 is used → 48.
The measuring device works with the Dynamic Host Configuration Protocol (DHCP), on
leaving the factory, i.e. the IP address of the measuring device is automatically assigned by
the automation system (DHCP server).
To establish a network connection via the service interface (CDI-RJ45): the "Default IP
address" DIP switch must be set to
ON
. The measuring device then has the fixed IP address:
192.168.1.212. This address can now be used to establish the network connection.
1. Via DIP switch 2, activate the default IP address 192.168.1.212: → 48.
2. Switch on the measuring device.
3. Connect to the computer using a cable → 74.