Commissioning
Proline Promass F 300 EtherNet/IP
146
Parameter
Prerequisite
Description
Selection
Factory setting
Totalizer operation mode
One of the following options is
selected in the
Assign process
variable
parameter
Totalizer
1 to n
submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Select totalizer calculation
mode.
• Net flow total
• Forward flow total
• Reverse flow total
Net flow total
Failure mode
One of the following options is
selected in the
Assign process
variable
parameter
Totalizer
1 to n
submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Define totalizer behavior in
alarm condition.
• Stop
• Actual value
• Last valid value
Stop
*
Visibility depends on order options or device settings