Proline Promass F 300 EtherNet/IP
Diagnostics and troubleshooting
205
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
520 I/O 1 to n hardware configuration invalid
1. Check I/O hardware
configuration
2. Replace wrong I/O module
3. Plug the module of double pulse
output on correct slot
0x1000276
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
528 Concentration settings faulty
1. Check concentration settings
2. Check input values e.g. pressure,
temperature
0x8000387
Status signal
S
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
529 Concentration settings faulty
1. Check concentration settings
2. Check input values e.g. pressure,
temperature
0x8000389
Status signal
S
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
537 Configuration
1. Check IP addresses in network
2. Change IP address
0x100014A
Status signal
F
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
594 Relay output simulation
Deactivate simulation switch output 0x20002BA
Status signal
C
Diagnostic behavior
Warning
12.8.4 Diagnostic of process
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
803 Current loop
1. Check wiring
2. Change I/O module
0x10000AD
Status signal
F
Diagnostic behavior
Alarm