Picomag IO-Link
Electrical connection
Hauser
17
Switch output configuration version
The switching behavior of IO1 and IO2 can be configured independently of one another.
PNP
NPN
1
2
3
4
B
A
L+
L-
A0033005
A
Plug (measuring device)
B
Socket (customer side)
L+ Supply v
L-
Supply voltage -
1
2
3
4
B
A
L+
L-
A0033006
A
Plug (measuring device)
B
Socket (customer side)
L+ Supply v
L-
Supply voltage -
The load is switched to high side L+. The maximum
load current is 250 mA. The output is protected
against short-circuiting and reverse polarity.
The load is switched to low side L-. The maximum load
current is 250 mA. The output is protected against
short-circuiting and reverse polarity.
1)
Positive Negative Positive (high side switch)
2)
Negative Positive Negative (low side switch)
Pulse output configuration version
Option only available for output 1 → 25
1
2
3
4
B
A
L+
L-
A0047257
3
Pulse output with PNP switching behavior
A
Plug (measuring device)
B
Socket (customer side)
L+ Supply v
L-
Supply voltage -
The load is switched to high side L+. The maximum load current is 250 mA. The output is
protected against short-circuiting and reverse polarity.