Picomag IO-Link
Appendix
Hauser
53
Designation
ISDU
(hex)
ISDU
(dec)
Size
(byte)
Data type Access
Value range
Factory setting
Range
limits
Totalizer Reset
0x016A
362
2
uint
w
cancel=0
reset=1
cancel
System Units
Unit Volumeflow
0x0226
550
2
uint
r/w
l/s=0
l/h=5
fl. oz/min=4
m³/h=1
l/min=2
Usgpm=3
l/min
Unit Volume
0x0227
551
2
uint
r/w
ml=0
USozf=1
l=2
m³=3
Usgal=4
ml
Unit Temperature
0x0228
552
2
uint
r/w
°C=0
°F=1
°C
Unit Conductivity
0x0229
553
2
uint
r/w
µS/cm=0
S/m=1
mS/cm=2
µS/cm=0
Unit Totalizer
0x016B
363
2
uint
r/w
USozf=1
l=2
m³=3
Usgal=4
kl=5
Ml=6
kUsg=7
m³
Sensor
Install. Direction
In relation to direction of arrow on
the device
0x015E
350
2
uint
r/w
forward=0
reverse=1
forward
Low Flow Cut Off
The flow rate below the selected
value is zero
Unit selection list from
Unit
Volumeflow
0x0160
352
4
float
r/w
0.4/0.75/1.2/5.0
l/min
0
10⁶
Damping
Volume flow damping via the PT1
element
Unit: s
0x01A4
420
4
float
r/w
0 s
0
100